Product Specifications OLD

The CONTRAfluranTM canister

The sensofluran sensor on a contrafluran canister

The SENSOfluran™ sensor2

  • The sensor is both a holder and a sensor. It clamps on to anesthesia machines and holds the canister.
  • The gas detector at the bottom of the sensor unit detects gas flow coming out of the canister as it fills.
  • An LED light panel indicates when the canister is full and needs to be changed.

SENSOfluran™ is intended to remove fluorinated halocarbons (such as sevoflurane, desflurane, and isoflurane).2 Use SENSOfluran PLUS to stay connected to the anesthetic gas scavenging system (AGSS) or if nitrous oxide will be used in the operating room.4

 

The sensor indicator system2

The green light on the SENSOfluran is on

GREEN LED (bottom):

Indicates the canister is ready to capture expired gas and has sufficient free capacity.

The green light and the first yellow light are on

YELLOW LED (second from bottom):

Indicates the capacity of the canister is diminishing. But the concentration of the anesthetic gas in the filtered exhausted air lies within the MAK accepted values. A canister change is recommended when the second yellow LED lights up.

Both yellow lights on the SENSOfluran are on

TWO YELLOW LEDs (middle two):

Indicates the maximum capacity of the filter is approaching. The canister should be replaced as soon as possible.

The red light on the SENSOfluran is on

RED LED (top):

Indicates the maximum capacity of the filter is reached. The canister must be replaced.

Document

Technical data2

CONTRAfluran™-Anesthetic gas canister

Temperature
  • Operation: +5 °C to +35 °C (41 °F to 95 °F)
  • Storage: -5 °C to +35 °C (23 °F to 95 °F)
Relative humidity
  • Operation: to 70%
  • Storage: to 70%
Storage capacity
  • Approx. 400 g (14 oz.)
Flow-resistance
  • ≤ 0.15 mbar
Height
  • 19 cm (7.5 in.)
Diameter
  • 12 cm (4.7 in.)
Weight
  • Approx. 1.000 g (0.353 oz.)
Volume
  • 2 L

SENSOfluran™-Fill level control unit

External material
  • Powder-coated aluminum
Temperature
  • +5 °C to +35 °C (41 °F to 95 °F)
Voltage
  • 100 to 240 V AC / 47 to 63 Hz
  • Output 6 V DC
  • Power consumption approx.: 6 W
Weight
  • Approx. 700 g (25 oz.)

Appropriate use

The CONTRAfluran™-Anesthetic gas canister is intended to remove fluorinated halocarbons (such as sevoflurane, desflurane, and isoflurane) in operation rooms, intensive care units, and mobile or ambulatory treatment areas, regardless of the type of exhaust device used. The CONTRAfluran™-Anesthetic gas canister is designed exclusively for these applications.

The SENSOfluran™-Fill level control unit must only be used in combination with the CONTRAfluran™-Anesthetic gas canister during sedation and operation of patients using volatile anesthetics.

CONTRAfluran™ and SENSOfluran™ are not intended for use with nitrous oxide or other anesthetic gases!

Please note that the canister and fill level control unit must only be used by qualified personnel.

 

For the safe and proper use of the devices mentioned herein, please refer to the complete Instructions for Use.

Summary of product characteristics for products mentioned on this website